MODEL 6201PORTABLE APPLIANCE TESTERINSTRUCTION MANUALKYORITSU ELECTRICAL INSTRUMENTSWORKS,LTD.
3.5 Explanation for indications3.5.1 LCD DisplayNote) Over range display: "OL" is displayed on the LCD.3.5.2 LED for threshold value indicat
―9―● General specification, measuring range and accuracyVoltage(VOLT) measurement of main power supplyMeasurement of Protective conductor resistance(R
―10―Reference test conditionUnless otherwise specified, this specification is dependent on followingcondition.(1) Ambient temperature: 23±5℃(2) Relat
―11―5.1 Visual inspectionBefore starting a measurement, user should undertake visual checks on the mains cord, case and that the correct type and rate
WARNING● When the voltage of mains power supply is 265V or more, "HI-V" isdisplayed on the LCD and buzzer sounds (discontinuous sounds). In
―13―(3) Insert Test Lead with safety alligator clip(M-7129) in to the E terminal of theinstrument, and contact the tip of the Test Lead with the meta
―14―6. Measuring method6.1 Earthed appliance TestThe purpose of the test carried out for Class 1 appliances is to check theinsulation resistance betwe
―15―(3) Connect Test Lead M-7129 to Rpe terminal and clip the metal part of DUTwith alligator clip.(4) Press the START switch when above preparatio
―16―6.2 Double Insulation TestThe Class 2 appliances have the indication of "DOUBLE INSULATION" or the symbol of F. Double insulation test
―17―CAUTION● When the terminal is open or the resistance value exceeds measuringrange, "OL" mark (over range display) appears on the LCD.● D
1.Safe testing ...12. Procedure of removing cover ...
―18―(1) Plug the extension leads adaptor M-7140 to the IEC socket of M-6201 connector block.(2) Connect the plug of the adaptor M-7140 for extension l
―19―When the fuse blows during use, please replace with new one according tobelow procedure.(1) Use a flat head screwdriver and turn it about 45°to l
―20―8. ServicesIf this instrument should fail to operate correctly, return it to your distributor.Please remember to give all the information possible
―21―9. Case and strap assemblyStrap belt and probe case can be attached to the instrument as below.Pass the strap belt down through the side panel of
DISTRIBUTORKyoritsu reserves the rights to change specifications or designsdescribed in this manual without notice and without obligations.92-1559 03-
―1―1. Safe testing! Electricity is dangerous and can cause injury and death. Always treat it with thegreatest of respectand care. If you are not quite
―2―DANGER● This instrument can be connected only to the commercial power of240V+10%-10%, 50Hz.● For safety reasons, only use the Test Leads designed t
―3―2. Procedure of removing coverModel 6201 have a dedicated cover to protect against an impact from theoutside and prevent the operation part, the LC
―4―3.1 Product summaryThe Model 6201 is a hand-held portable appliance tester, performing fourfunctions to ensure the Safety of Class I and Class II a
(1) Test socket (2) Terminal block(6)LCD (7) LED for test result (8)Start switch(9)Null/250V-500V switch(10) Range switch(11) (3) (4) (5) Termi
―6―(1)Test socketInsert the mains plug of DUT to this socket for the polarity test of protectiveconductor resistance, insulation resistance and Extens
(11)FuseProtected by a fuse of 600V/500mA ceramic fuse (F type Φ6.3x32mm).User can replace this fuse.(12)Mains cord(AU) M-7123This mains cord can be c