Kyoritsu 2009R Instrukcja Użytkownika Strona 28

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 33
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 27
̶
26
̶
 ② Slide the terminal cover to the right to disclose the OUTPUT
terminal and insert the output plug into the terminal. Make
connection to the recording device.
 ③ When the plug is inserted into the output terminal, auto-range
function is cleared.
  Set the range depending on the state of Data Hold Switch.
Data Hold Switch OFF 400A range
Data Hold Switch ON 2000A range
Note: After measurement, be sure
to return Data Hold Switch to OFF
position.
 ④ Set the function selector switch to the desired position (ACA or
DCA) and follow appropriate measurement instructions.
NOTE
 ● During current measurement, keep the transformer jaws fully
closed. Otherwise, accurate measurement cannot be made.
The maximum measurable conductor size is approx. 55mm in
diameter.
 ● Zero adjustment is not necessary on AC cur
rent range.
 ●
On DC current range, the Zero Adjust/Reset button may not
completely zero adjust the output voltage from the OUTPUT
terminal. In this case, make zero adjustment on the recording device.
 ● Connecting the plug to the OUTPUT terminal disables the Auto-
power-off function. The function is enabled on removing the plug
from the terminal.
Przeglądanie stron 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33

Komentarze do niniejszej Instrukcji

Brak uwag