Temperature Measurement TEMP
Range Resolution Accuracy
Input Protective
Voltage
-50 ~ 600°C 0.1°C 2+2°C
1000V rms
-58~ 999.9°F
-58~ 1112°F
0.1°F
1°F
2+3.6°F
2+3°F
Use optional Temperature Probe: Thermocouple Type K
Capacitor Measurement
Range Resolution Accuracy
Input Protective
Voltage
10nF 0.01nF 2+10
1000V rms
100nF 0.1nF
2+51μF 0.001μF
10μF 0.01μF
100μF 0.1μF
3+5
1000μF 1μF
Accuracy is specied after zero adjustment at 10n to 1μF (Capacitance).
Frequency Measurement Hz
AC Coupling, Maximum Reading 9999
Range Resolution Accuracy Input Voltage
10.00 to 99.99Hz 0.01Hz
0.02+1
0.2 to 600Vrms
90.0 to 999.9Hz 0.1Hz
0.900 to 9.999kHz 0.001kHz 0.4 to 600Vrms
9.00 to 99.99kHz 0.01kHz 0.8 to 100Vrms