1) Display “Ac” by using the LIGHT key or SHIFT key.
HOLD
RMS
MEAN
2) Select the detection method by using the ▲ (RANGE) key or ▼ (REL) key.
3) Press the HOLD key to save the setting.
4) “SEt” appears and then “Ac.”
<Default value of LOGGING interval> (KEW1052 only)
Set a default value of the saving interval during LOGGING mode.
1) Display “L.Int” by using the LIGHT key or SHIFT key.
The default setting is 1 sec.
HOLD
2) Select the saving interval by using the ▲ (RANGE) key or ▼ (REL) key.
3) Press the HOLD key to save the setting. “SEt” appears and then “L.Int.”
Settings of saving interval
1, 2, 5, 10, 30, 60, 600, 1800 sec
<Sound on/off setting>
Set the sound on/off (beep of buzzer)
Even if the user sets the sound off, it goes off at the following points.
checking continuity
alarm for over-input
alarm for auto power-off
1) Pressing the LIGHT key or SHIFT key shows “bEEP” on the display.
Set on/off on the sub-display. Default is ON.
HOLD
: ON
: OFF
2) Select on/off by using the ▲ (RANGE) key or ▼ (REL) key.
3) Press the HOLD key to save the setting.
“SEt” appears and then “bEEP.”