Kyoritsu 6305 Instruction Manual Instrukcja Użytkownika Strona 31

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 30
KEW6305 4.4
Setting 02” Voltage rang
Selecting a measurement range so that the estimated inputs will be close to the full scale value is
recommended to obtain accurate results. Recommended range selections are: 150V range for rated
voltages between 100 - 120V, 300V range for 200 - 240V and 600V range for 400 – 440V.
Setting item 150V / 300V / 600V
Default value (or after system reset) 300V
1 Use the Cursor key on the selection screen, and select “Setting 02”.
2 Press the ENTER key to get the instrument into setting change mode.
3 Present setting (or default value: 300V) blinks. Select the appropriate voltage range with the Cursor
key, and then press the ENTER key after making necessary change.
Przeglądanie stron 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 137 138

Komentarze do niniejszej Instrukcji

Brak uwag